eCHEM Chlordioxid up to 2 mg/l ClO2

Input data
Cable specification black PUR (halogen free), shielded, M12 plug
Temperature compensation Automatic through integrated temperature sensor, temperature changes <5 °C/h
Measured variables Chlorine Dioxide
Measuring range 2 ppm, 20 ppm
Function and system design
Measuring method Membrane-covered, amperometric 2-electrode system
Measuring method Amperometry
Auxiliary power
Electrical connection 8-pin M12 plug
Power supply 12-24 V
Output data
Output signal RS-485, Modbus RTU
Accuracy Measuring range 2 ppm: at 0.4 ppm & 1.6 ppm < 1 %

Measuring range 20 ppm: at 1.5 pm < 0.1 %

Data interface RS-485, Modbus RTU
Environmental conditions
Compressive strength 1.0 bar, no pressure shocks or vibrations
Compressive strength 1.0 bar, no pressure shocks or vibrations
Process conditions
Process temperature +5…+50 °C
pH range 1.0 bar, no pressure shocks or vibrations
pH range pH 2…11
Performance characteristics
Response time T90: approx. 1 min
Running-in period Approx. 1 h prior to initial operation
Cross influences Cl2: does not interfere, O3: is measured with 25 higher sensitivity than ClO2
Calibration method On the meter by means of analytical determination
Maintenance interval Regular monitoring of the measurement signal at least once a week.
Structural design
Materials PVC-U, stainless steel 1.4571
Certificates and approvals
Work certificates Yes
CE Manufacturer’s declaration 2014/30/EC EMC Directive (EN 61326-1:2013)
Scope of delivery
value only lead Manual Yes
Test certificate Yes
Accessories
Fittings Flow unit
Cables Extension cables: 0.3 m, 2 m, 10 m, 25 m
Measuring transducer TriBox3, TriBox mini, HS100, TriBox Flex
Validation/calibration DPD method