| Auxiliary power | |
|---|---|
| Interface digital | RS-232 (TriOS) |
| Interface analog | 4…20 mA, 0…5 V |
| Power supply | 12-24 VDC (± 10 %) |
| Input data | |
| Turbidity compensation | No |
| Data logger | No |
| T100 response time | < 10 s |
| Measurement interval | > 5 s |
| Measurement accuracy | enviroFlu-HC 500: 0.3 ppb |
| Measurement accuracy | enviroFlu-HC 5000: 0.5 ppb |
| Function and system design | |
| Analytical methods | Fluorescence |
| Light source | Miniature Xenon-flash lamp with interference filter (excitation wavelength 254 nm). Output energy controlled by internal reference diode. |
| Output data | |
| Power consumption | < 3.5 W |
| Performance characteristics | |
| Maintenance effort | Typically < 0.5 h/month |
| Calibration/maintenance interval | 24 months |
| System compatibility | Analog Out (0…5 VDC, 4…20 mA) |
| Guarantee | 1 year (EU: 2 years) |
| Installation | |
| Max. pressure with SubConn | 30 bar |
| Max. pressure with fixed cable | 3 bar |
| Max. pressure in flow cell | 1 bar, 2-4 L/min |
| Protection type | IP68 |
| Structural design | |
| Housing material | Stainless steel (1.4571/1.4404) or titanium (3.7035) |
| Weight stainless steel | ~ 2.7 kg |
| Weight titanium | ~ 1.9 kg |