| Dimensions | |
|---|---|
| Length | 205 mm |
| Diameter | 25 |
| Input data | |
| Measured variables | Free chlorine with reduced pH dependency |
| Temperature compensation | Automatic through integrated temperature sensor Pt100 |
| Cable specification | black PUR (halogen free), shielded, M12 plug |
| Function and system design | |
| Measuring method | Amperometry |
| Measurement principle | Membrane-covered, potentiostatic amperometric 3-electrode system |
| Auxiliary power | |
| Electrical connection | 8-pin M12 plug |
| Power supply | 12-24 V (± 10 %) |
| Output data | |
| Data interface | RS-485, Modbus RTU |
| Output signal | RS-485, Modbus RTU |
| Accuracy | Measuring range 2 mg/L: at 0.4 mg/L & 1.6 ppm < 1 % Measuring range 20 mg/L: at 4 mg/L < 1 %, at 16 mg/L W37< 3 % |
| Process conditions | |
| Process temperature | 0…+45 °C (no ice crystals in the test water) |
| Process pressure | 3 bar, no pressure shocks or vibrations |
| Flow rate | Approx. 15…30 L/h in FLC-3, minimum flow dependence exists |
| pH range | pH 4 … pH 9, reduced pH dependence |
| Conductivity | 10 µS/cm…50 mS/cm (sea water) |
| Performance characteristics | |
| Response time | T90: approx. 2 min |
| Running-in period | Approx. 2 h prior to initial operation |
| Drift | Approx. -1 % per month |
| Cross influences | Combined chlorine increases measured value |
| Calibration method | Determination of chlorine with DPD-1 method |
| Certificates and approvals | |
| CE Manufacturer’s declaration | 2014/30/EC EMC Directive (EN 61326-1:2013) |
| Work certificates | Yes |
| Scope of delivery | |
| Manual | Yes |
| CE Manufacturer’s declaration | Yes |
| Test certificate | Yes |
| Accessories | |
| Fittings | Flow unit |
| Cables | Extension cables: 0.3 m, 2 m, 10 m, 25 m |
| Measuring transducer | TriBox3, TriBox Mini, HS100, TriBox flex |
| Validation/calibration | DPD method |