| Auxiliary power | |
|---|---|
| Electrical connection | 8-pin M12 plug |
| Power supply | 12-24 V |
| Function and system design | |
| Measurement principle | 90° scattered light method |
| Measuring method | Nephelometry |
| Input data | |
| Cable specification | black PUR (halogen free), shielded, M12 plug |
| Measuring range | 0…30 g/L |
| Measurement wavelength | 860 nm |
| Output data | |
| Output signal | RS-485, Modbus RTU |
| Accuracy | ± 3 % FS |
| Data interface | RS-485, Modbus RTU |
| Power consumption | 3 W |
| Process conditions | |
| Process temperature | 0…+60 °C |
| Process pressure | 4 |
| Performance characteristics | |
| Repeatability | 98 % |
| Response time | 90 % of the value in 5 seconds |
| Calibration method | On the meter by means of analytical multipoint determination |
| Installation | |
| Protection type | IP68 |
| Structural design | |
| Thread | 1” GAS |
| Materials | Stainless steel 1.4401 casing, window with epoxy glue, Viton® O-ring |
| Certificates and approvals | |
| CE Manufacturer’s declaration | 2014/30/EC EMC Directive (EN 61326-1:2013) |
| Work certificates | Yes |
| Scope of delivery | |
| value only lead Manual | Yes |
| CE Manufacturer’s declaration | Yes |
| Test certificate | Yes |
| Accessories | |
| Cables | Extension cables: 0.3 m, 2 m, 10 m, 25 m |
| Measuring transducer | TriBox3, TriBox mini, HS100,TriBox Flex |